When operate temperature humidity chamber,users may encounter such situation, humidity suddenly showed 100% RH, but do not know what the reason? Today I will address this issue in detail for your reference.

Step 1: Check if the gauze is hung correctly on the humidity sensor, which is close to the sink in internal chamber. One end of the gauze should be soaked in water, the other end should be hung on humidity sensor. Some customers often hang the gauze in wrong position, thus cause the machine failure. Therefore please carefully identify the sensors before operation.


Step 2: If the hanging mode is correct, humidity still shows 100%RH. Please check whether the wet ball gauze has yellowish phenomenon, or use for too long, too much minerals are accumulated on the gauze, which may reduce the water-absorbing performance of the gauze. So the detected humidity is not accurate, resulting in high humidity of the illusion; if so please replace to new gauze. If not, check water level of water sink. If water level is low, take the following steps:

A. Observe the water tank side of the water level box whether there is water, if not, there may be damage to the floating ball.

B. Electromagnetic pump damage, please measure its resistance with a multimeter, if open circuit, it means this pump is damaged.

C. Gap replenishment timer is damaged, can not be normal replenishment.

Step 3: The controller setting may be wrong, some customers may set humidity at very low temperature (minus) or at high temperature (above 100 degrees celsius), this situation will also cause control Misalignment and instability. Humidity can't be controlled below 0 degree,it's easy to ice; during high temperature, evaporation is too fast, so it's very difficult to control humidity. Therefore, please carefully read the instructions, regulate the operation of machinery, extend equipment life.

Dongguan Yuanyao Electronics Technology Co.,Ltd is a 8-year-old company specializing in the manufacture of temperature humidity chamber and other environmental chambers.Our customers range from factories, medical facilities,R&D centers, universities to accredited certification laboratories.We are committed to provide good quality products and excellent service to our customers.Your case may be unique, but we will certainly be able to resolve it in the best possible way.

If you are interested in our temperature humidity chamber, please write to robinwang00 for more details.

3 条评论:

  1. Your blog is very helpful for more information about Environmental climate chamber. Best company of Environmental test chamber manufacturers in India.

    回复删除
  2. The Humidity Chamber is an essential piece of equipment designed to control and maintain specific environmental conditions for precise testing purposes. It creates a controlled atmosphere with regulated humidity levels, allowing researchers, scientists, and manufacturers to simulate real-world conditions accurately. By adjusting temperature and humidity settings, the chamber can replicate a wide range of environmental scenarios, including tropical, arid, or humid conditions. This controlled environment is crucial for conducting various experiments, tests, and studies across multiple industries, such as electronics, pharmaceuticals, and materials science. The Humidity Chamber ensures the reliability and reproducibility of test results, facilitating product development, quality assurance, and research endeavors.

    回复删除
  3. Thank you for this insightful article on troubleshooting temperature humidity chambers! It’s great to see such detailed steps laid out for users who might encounter the frustrating issue of suddenly reading 100% RH. I appreciate the emphasis on checking the gauze position on the humidity sensor; it's such a simple yet crucial step that can easily be overlooked. The tips on gauze maintenance and the potential issues with the water level are particularly helpful as they provide a clear path for diagnosing the problem.